[2] For in vitro diagnostic (i.e. outside of the body) use only.
[3] With AgaMatrix WaveSense JAZZ Test Strips and AgaMatrix
Control Solution only. Do not use other brands of test strips and
control solution with this meter. This may produce inaccurate
results.
[4] For use at home (over the counter (OTC)) by persons with
diabetes, or in a clinical setting by healthcare professionals, as
an aid to monitor the effectiveness of diabetes control.
Do not use the AgaMatrix WaveSense JAZZ System for:
[1] The diagnosis of diabetes.
[2] Testing the glucose levels of neonates (children under 4
weeks).
[3] Testing glucose levels of arterial or venous blood.
[4] Testing glucose from sites other than the fingertip, palm
(at the base of the thumb), or forearm.
Possible test sites:
The AgaMatrix WaveSense JAZZ System requires only 0.5 μL
of blood to perform a glucose test. For best results, test using
blood from the tip of any finger. The palm (at the base of the
thumb) and forearm are alternate, but less accurate test sites.
Only select soft and fleshy areas to lance. Avoid lancing any
bony areas, obvious veins and moles.