AgaMatrix Jazz Specifications Page 56

  • Download
  • Add to my manuals
  • Print
  • Page
    / 66
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 55
54
Battery Icon:
Battery #2 has died during test.
Replace battery #2 immediately before retesting.
Troubleshooting Situation #1: Meter does not enter the test
mode after inserting an AgaMatrix WaveSense JAZZ Test Strip.
[1] CAUSE: The batteries have insufficient power. ACTION:
Replace the batteries, set time and date, and test.
[2] CAUSE: The batteries are installed incorrectly or there are
no batteries in the meter. ACTION: Check that both batteries
are installed correctly, with the positive (+) sign facing up
towards you.
[3] CAUSE: AgaMatrix WaveSense JAZZ Test Strip has been
inserted upside down, wrong end in, or incompletely inserted
into the meter. ACTION: Insert the AgaMatrix WaveSense JAZZ
Test Strip with the black side up and the contact bars of the test
strip in the meter. Ensure that the test strip is fully inserted.
[4] CAUSE: Defective meter or defective AgaMatrix
WaveSense JAZZ Test Strips. ACTION: Call customer care.
[5] CAUSE: Blood or foreign objects introduced into the
AgaMatrix WaveSense JAZZ Meter test strip port. ACTION:
Please refer to chapter 22. If problem persists, call customer
care and have the meter serial number available (located on
back of the meter).
Troubleshooting Situation #2: The glucose test sequence
does not start after applying the blood sample.
[1] CAUSE: Defective test strip. ACTION: Repeat the test